ASG150 Insulation Test Signal Generator
● Generate Fault Location Test Signal
● Detection of L1&L2 Disconnection Status
● LED Indicator
● CAN bus Communication
● IEC 61557-9-2007

PRODUCTS

  • Features

    Generate Fault Location Test Signal


    ● Location Current: <1mA


    ● Location Voltage: <25V

    1.jpg


    LED Indicator


    ● On-off (connection status of devices)


    ● Communication Status


    ● L1&L2 Disconnection Status


    ● Insulation Fault

    2.jpg



    CAN bus Communication


    ● Customized Protocol


    ● Support Multiple Master

    3.jpg


    Wiring Instruction

    4.jpg


    Typical Connection

    6.jpg



  • Auxiliary power
    VoltageDC24V(±25%)
    The maximum power consumption≤2VA
    System be monitoredRated voltage0~ 242V AC
    Rated frequency45 ~ 60Hz
    Location signalsLocation voltage<25V
    Location current<1mA
    CommunicationBusCAN
    Protocol Custom protocol

  • External dimensions of ASG100 test signal generator (unit: mm)


                                                

    Front view                                                                                            Side view